S1100 ICU Ventilator

Category:

Description

Application

The ventilator makes a good performance in operation room, ICU department and emergency treatment. It used to assist or replace the spontaneous breathing for adult pediatric and neonatal more than 2kg.Due to the flexible configuration, good quality and competitive price, S1100 has soon become the superstar of market.

Features

1.15” TFT LCD screen displays the Ventilation parameters, Alarming information and oscillograms, make every operation more easily.

Multiple ventilation modes, suitable for ICU, emergency department and operation room etc:(PRVC,APRV,DUOLEVEL,V-SIMV,P-SIMV,IPPV,A/C,PCV,PSV,SPONT/CPAP, SIGH, MANUAL)

2.5 oscillograms for your choice, 3 of them can be displayed on the screen at the same time.

3.Humidifier can heat and wet breathing gas,makes it comfortable for patient to breathe

4.Rapid oxygen supply, automatically offer high flow rate oxygen within two minutes.

5.Nebulization make medicine into small liquid and particle, easier and quicker for patient to breathe in.

6.High temperature resistance breathing circuit is reusable and anti-pollution.

 

Safety

1.14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting.

2.Built-in oxygen concentration sensor, ensure stable precision of oxygen concentration

3.Easy to move with four casters, easy to stop with two brakes

4.Separate design of electronic circuit and gas flow rate keep safe running of ventilator.

5.Compact long life internal battery can provide emergency power, avoid risk of patient

6.Self-check before operation, eliminate system mistake

Technical Specifications

NAME PARAMETER
Ventilation modes PRVC,APRV,DUOLEVEL,V-SIMV,P-SIMV,IPPV,A/C,PCV,PSV,SPONT/CPAP, SIGH, MANUAL
Ventilator parameter ranges
Tidal volume(Vt) 0~2000mL
Frequency (Freq) 1min~100min
Oxygen concentration 21%-100%
I/E 4:1~1:8
PEEP 0cmH2O~ 40 cmH2O
Pressure Triggering Sensitivity (Ptr) -20cmH2O~20cmH2O (Based on PEEP)
Flow Triggering Sensitivity (Ftr) 0.5L/min~30L/min
Pressure Control 5cmH2O~80cmH2O
Pressure Support 0cmH2O~80cmH2O
SIGH 0(off) 1/100~5/100
Apnea Ventilation OFF, 5s~60s
Pressure Limit 20cmH2O~100cmH2O
Monitored Parameters
Frequency (Freq) 0 /min~ 100 /min
Tidal volume(Vt) 0 mL~ 2500 mL
MV 0L/min~99L/min
Airway pressure 0cmH2O~100cmH2O
Dynamic Lung Compliance Testing 1mL/cmH2O~1000mL/cmH2O
Oxygen Concentration 15%-100%
Alarm and protection  
The AC power failure alarm Power failure or no connection
Internal battery backup low voltage alarm ≤11.3±0.3V
No tidal volume No tidal volume within 6s
High Minute Volume alarm

Low Minute Volume alarm

5L/min-99L/min

1L/min-30L/min

High Airway pressure alarm

Low Airway pressure alarm

High oxygen concentration alarm

Low oxygen concentration alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

19%-100%

18%-99%

Continuous pressure alarm (PEEP+1.5cmH2O) over 16s
Suffocation warning

Fan error

Oxygen deficit

5-60s

Show on screen

Show on screen

The maximum limited pressure <12.5 kPa
Working conditions
Gas source O2,AIR
Pressure 280kPa-600kPa
Voltage 220V±22V
Power frequency 50Hz±1Hz
Input power 900VA(with air compressor)

250VA(without air compressor)

Oscillogram
P-T Pressure Time
F-T Flow-Time
V-T Volume Time
P-V Loop Pressure-Volume Loop
F-V Loop Flow Volume Loop
P-F Loop Pressure-Flow Loop