Description
Features
-10.4 inch color LCD touch screen with knob (self-adaptive brightness adjustment)
-Night and day model switches
-Low lightness far infrared ceramic heating lamp
-Ambient noise and light level display
-Hand free alarms silence
-Integrated SpO2 function (optional)
-Thermal conduction gel mattress
-Far infrared ceramic heating tube
-Self-adaptive Brightness Adjustment Touch Screen
-Integrated weight scale and resuscitation (optional)
-With storage drawer as standard configuration
Standard Configurations
PARAMETER | SPECIFICATION |
Power Requirements | Customer Specified |
Power consumption:700VA (Maximum) | |
Heater capacity | 550W |
Dimensions | 84(W)x110(D)x178-218(H)cm (Mattress surface:80-120cm) |
Weight | 100KG |
Temperature Control | Servo control (manual control also available) |
Skin Temperature | Setting range:34.0-38.0 °C |
Display range:30.0-42.0 °C | |
Heater Output | Setting range:0-100% (in 5%increments) |
Indication Range:0-100 (in 10 levels) | |
Timer | Time display range:0-60 minutes 00 seconds (in 1 second increments) |
APGAR mode/CPR mode switchable | |
Lighting | LED lamp Adjustable illumination intensity 0-1200lux |
Mattress platform angle | Inclination angle (back and forth):+13 °~-13 ° |
Optional Functions
PARAMETER | SPECIFICATONS |
Height Adjustment | Up to 40cm |
Weight Scale | |
Weight range | 300g to 8kg |
Weight display resolution | 1g |
Weight accuracy | +/-10g |
Pulse Oximeter Unit | |
Display range | 1-100% SpO2 |
SpO2 alarm setting range | Upper limit:50-99% (in 1% increments) |
Lower limit:45-95% (in 1% increments) | |
Pulse rate display range | 25-240bpm |
Pulse rate alarm setting range | Upper limit:80-240bpm, OFF |
Lower limit:35-180bpm, OFF (in 5bpm increments) | |
Alarms | SpO2 high/low limit, pulse rate high/low limit, SpO2 sensor |
Resuscitation unit | |
Manometer | Pressure indication:-2~8kPa (-20~80cmH2O) |
Suction unit | Suction pressure control range:0~26.7kPa (0-200mmHg) |
O2 Blender | Oxygen concentration control range:21~100% |
Oxygen flow rate control range:0~15L/min | |
Resusci-flow unit | PMAX setting range:19~58kPa (20~60cmH2O) |
Default setting range:39kPa(40cmH2O) | |
Storage/Shipping Information | |
Temperature | -25°to 60°C |
Humidity | 0-95%RH |