ARI-9800D Mammography System

Description

Features

  1. The 9800D high frequency molybdenum target mammography machine are featured with very small focus, high resolution, low dosage and micro-computer controlled.
  2. It is a high-performance breast screening machine for iconography examination of mammary diseases and with high sensitivity for mammary benign lumps, nodes, calcified points and early detection of malignant.
  3. Easy to use user interface design with finger touch operation panel, fast reachable buttons and foot switch.
  4. Streamlined compressor plate provides the intact breast images.

High quality carbon fiber catridge table with low X-ray absorption that reduces scattered X-ray.

  1. The 80KHz high frequency generator designed with advanced IGBT high-frequency inversion technology and quasi-resonant pressure boosting technology.
  2. It provides high efficiency, stable voltage, pure output spectral line that produces high-definition images.
  3. Reliable and durable mechanical design provides:

Electrical up/down moving smoothly and quiet.

360 free rotation C-arm provides imaging flexibility.

  1. Main Parameters

Two different X-ray tubes to choose:

— Hangzhou LR01 tube from China

— IAE C339V tube from Italy

  1. X-ray Generator

Complete solid-state high-frequency high-voltage generator.

Generator Type: High Frequency Inverter 80kHz

Input Power: Single phase 220VAC, 50/60Hz

Radiographic Ratings: Large Focal Point 20-35kV/10-510mAs

                    Small Focal Point 20-35kV/10-100mAs

Power Rating: 6kW

  1. X-ray Tube

1)     Hangzhou LR01 tube from China

Focal Spot Size: Dual Focus 0.2 / 0.4mm

Target Material:Molybdenum (Mo)

Port Material:Beryllium (Be)

High-speed anode drive: 2800/1000rpm

Target angle: 12°/12°

Anode Heat Storage: 100KJ (150kHU)

Anode Cooling:Air cooling

Filtration: Mo(0.03mm),Al(0.5mm)

2)     IAE C339V tube from Italy

Focal Spot Size: Dual Focus 0.1/0.3mm

Target Material: Molybdenum (Mo)

Port Material: Beryllium (Be)

High-speed anode drive: 2800/1000rpm

Target angle: 10°/16°

Anode Heat Storage: 210kJ (300kHU)

Anode Cooling: Air cooling

Filtration: Mo(0.03mm), Al(0.5mm)

  1. Radiographic Stand

C-ARM: Vertical Movement: 590mm

Center of electric rotating C-arm, automatic return function by one key

Automatically released after the exposure pressure settings display

Compression flexible stepless speed.

Max. pressure: 200N

Max. travel: 150mm

Rotations Deg:+90°~-90°

SID: 650mm

  1. Cassette Image Receptor

18*24cm (24cm*30cm optional)

Bucky Device: 18*24cm Bucky drive mechanism,(24cm*30cm optional)

Grid Ratio: 5:1, 30 Line/cm

  1. Line Voltage

220Vac@25A, Single phase

  1. Weight and Size

240kg(Gross weight: 340kg)

L*W×H: 940mm*560mm*2000mm