Introduction
Our modular pationt monitor can achieve modularization for function expanding with the help of integration plug-in technology, which can meet the requirement of updating and the monitors in all departments in the hospital. With various function, this machine can satisfy the convenient operation and the selection of all data of monitor in all departments in the hospitals.
Features
Integrated Plug in Modular, 12.1” colorful and clear TFT display
Standard 6 parameters and optional single plug-in parameter box
Touch screen available; able to be operated by keyboard or optional mouse
With durable trim knob and backlight silicone buttons
Simultaneously display at least 10 parameters, 13 kinds of arrhythmia analysis
ECG Waveforms of 7-lead displayed simultaneously.
Assist to judge the heart activity directly
15 kinds of drug dose calculations
Anti-electrosurgical, anti-defibrillation, suitable for monitoring critical care
Trend interface, OxyCRG interface, BigFont interface
Built-in detachable rechargeable lithium battery, battery life 4.5 hours
1000 hours trend data and trend graphs, 48 hours full waveform reviewing
Power failure data store
High capacity of information storage
NIBP dual overpressure protection
Wired/wireless connection, with system supporting 3 display screens
Supporting SD card, 4 USB interfaces
Omni-directional visible design; 3 levels of audible, visual and audio alarm; dual alarm lights for physiological and technical alarm
Technical specification
ECG
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection: I, II, III, avR, avL, avF, V,
Waveform: 2 ch
Lead mode: 3 Leads (R, L, F or RA, LA, LL)
Lead selection: I, II, III,
Waveform: 1 ch
Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto
Heart Rate
Range
-Adult: 15 ~ 300 bpm
-Neo/Ped: 15 ~ 350 bpm
Accuracy: ± 1% or ± 1bpm, which great
Resolution: 1 bpm
Sensitivity: > 200 (uV P-P)
Differential Input Impedance: > 5 MΩ
CMRR
-Monitor: > 105 dB
-Operation: > 105 dB
-Diagnosis: > 85 dB
Electrode offset potential: ±300mV
Leakage Current: < 10 uA
Baseline Recovery: < 3 S after Defi.
ECG Signal Range: ±8 m V (Vp-p)
Bandwidth
-Surgery: 1 ~ 15 Hz
-Monitor: 0.5 ~ 35 Hz
-Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy: ±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available
Review: Available
Respiration
Method: Impedance between R-F (RA-LL)
Differential Input Impedance: >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 – 2.5 KΩ
Bandwidth: 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
-Adult: 0 ~ 120 rpm
-Neo/Ped: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy: ±2 rpm
Apnea Alarm: 10 ~ 40 S
NIBP
Method: Oscillometric
Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,
120, 180, 240,480 (Min)
Measuring Period in STAT Mode: 5 Min
Pulse Rate Range: 40 ~ 240 bpm
Alarm Type: SYS, DIA, MEAN
Measuring and alarm range
Adult Mode
-SYS: 40 ~ 270 mmHg
-DIA: 10 ~ 215 mmHg
-MEAN: 20 ~ 235 mmHg
Pediatric Mode
-SYS: 40 ~ 200 mmHg
-DIA: 10 ~ 150 mmHg
-MEAN: 20 ~ 165 mmHg
Neonatal Mode
-SYS: 40 ~ 135 mmHg
-DIA: 10 ~ 100 mmHg
-MEAN: 20 ~ 110 mmHg
Resolution Pressure: 1mmHg
Accuracy Pressure Maximum Mean error: ±5mmHg
Maximum Standard deviation: ±8mmHg
Overpressure Protection
-Adult Mode: 297±3 mmHg
-Pediatric Mode: 240±3 mmHg
-Neonatal Mode: 147±3 mmHg
SpO2
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified
Actualization interval: about 1 Sec.
Alarm Delay: 10 Sec.
Pulse Rate
Measuring and Alarm Range: 20~300bpm
Resolution: 1bpm
Accuracy: ±2bpm
Temperature
Channel: 2
Measuring and Alarm Range: 0 ~ 50 °C
Resolution: 0.1°C
Accuracy: ±0.1°C
Actualization interval: about 1 Sec.
Average Time Constant:<10 Sec.
Configuration
Standard: ECG, RESP, NIBP, SPO2, PR, TEMP