Description
Technical Specifications
| PARAMETER | SPECIFICATION | 
| Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant | 
| Scope of Application | Adult, child | 
| Gas Source | O2: 0.27–0.55Mpa;
 N2O: 0.27–0.55MPa.  | 
| Flow Meter | O2: 0.1–1.0L/min; 1.1–10L/min
 N20: 0.1–1.0L/min; 1.1–10L/min  | 
| O2, N2O Linkage and N2O
 Stopper  | 
When using nitrous oxide, oxygen concentration>25%;
 When the oxygen pressure < 0.2 MPa, the flow of nitrous oxide would be cut off.  | 
| Flow Rate of Rapid Oxygen Supply | 25–75L/min | 
| Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure < 0.2MPa | 
| Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
 Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.  | 
| Respiratory Circuit | Working mode: all-close, semi-close, semi-open;
 Releasing pressure: 0.5-7kPa.  | 
| Respiratory Bellows | Bellows for adults, bellows for children, tidal volume range:0-1500 ml. | 
| VENTILATOR | |
| Display Mode | 10.4” Touch Screen, High-definition TFT color LCD. | 
| Ventilation Mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH (IMV, VCV, PCV) | 
| Ventilation Parameters | Tidal volume           50–1500ml
 Rate 1–99bpm SIMV rate 2–20bpm I:E 2:1–1:8 (4:1–1:8) Inspiratory trigger pressure -1.0–2.0kPa PEEP 0–2.0kPa Pressure range 1.0–6.0kPa SIGH 1.5 times the inspiratory time (70-150 adjustable) Inspiratory platform 0–50%  | 
| Parameters for Ventilation
 Monitoring  | 
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure – time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform
 (autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop)  | 
| Monitoring of Oxygen
 Concentration  | 
21%–100% | 
| SECURITY ALARM SYSTEM | |
| Oxygen Concentration alarm | Upper limit setting range   21%–100%;
 Low limit setting range 10%–80%.  | 
| Airway Pressure Alarm | Upper limit setting range   0.3–6.0kPa;
 Low limit setting range 0.2–5.0kPa.  | 
| Per-minute Ventilation Volume Alarm | Upper limit setting range   3.0–30L/min;
 Low limit setting range 1.0–10L/min.  | 
| Sustained High-pressure
 Alarm  | 
It will give alarm when stress have consistently been higher than 2.5 kPa | 
| Suffocation Alarm | It will give sound and light alarm if there is no tidal volume input for 15 seconds | 
| POWER ALARM | |
| Power | AC 220V±10%, 50Hz±1Hz, (Storage battery, UPS) | 
| Mechanical Arms (End-expiratory CO2 monitoring) | |
| Note: Items within brackets are optional. | |