Description
echnical Specifications
PARAMETER | SPECIFICATION |
Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of Application | Adult, child |
Gas Source | O2: 0.27–0.55Mpa;
N2O: 0.27–0.55MPa. |
Flow Meter | O2: 0.1–10L/min;
N2O: 0.1–10L/min |
O2, N2O Linkage and N2O
Stopper |
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure < 0.2MPa, the flow of nitrous oxide would be cut off. |
Flow Rate of Rapid Oxygen Supply | 25–75L/min |
Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
Respiratory Circuit | Working mode: all-close, semi-close, semi-open;
Releasing pressure: 0.5-7kPa. |
Respiratory Bellows | Bellows for adults, bellows for children, tidal volume range 0-1500ml |
VENTILATOR | |
Display Mode | High-definition 5.7” LCD screen display |
Ventilation Mode | IPPV, SIPPV, SIMV, PEE, MANUAL, SIGH |
Ventilation Parameters | Tidal volume 50–1500ml
Rate 2–99bpm SIMV rate 2–20bpm I:E 2:1–1:8 Inspiratory trigger pressure -1.0–2.0kPa PEEP 0–2.0kPa Pressure range 1.0–6.0kPa SIGH 1.5 times the inspiratory time |
Parameters for Ventilation
Monitoring |
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure – time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure. |
Monitoring of Oxygen
Concentration |
(21%–100%) |
SECURITY ALARM SYSTEM | |
Oxygen Concentration alarm | Upper limit setting range (50%–100%)
Low limit setting range (15%–50%) |
Airway pressure alarm | Upper limit setting range 1.0–6.0kPa;
Low limit setting range 0.4–2.0kPa. |
Per-minute ventilation volume alarm | Upper limit setting range 3.0–30L/min;
Low limit setting range 1.0–10L/min. |
Sustained high-pressure
alarm |
It will give alarm when stress have consistently been higher than 2.5 kPa |
Suffocation alarm | It will give sound and light alarm if there is no tidal volume input for 15 seconds |
POWER ALARM | |
Power | AC 220V±10%, 50Hz±1Hz, (Storage battery, UPS) |
Note: Items within brackets are optional. |