AR-321 Anesthesia Machine

Category:

Description

Technical Specifications

PARAMETER SPECIFICATION
Main Body High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of Application Adult, child
Gas Source O2 0.27–0.55MPa
Flow Meter O2: 0.1–1.0L/min; 1.1–10L/min
Flow Rate of Rapid Supply 25–75L/min
Low Oxygen Pressure Alarm There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.

Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.

Respiratory Circuit Working mode: all-close, semi-close, semi-open

Releasing pressure: 0.5-7kPa.

Respiratory Bellows Bellows for adults, bellows for children, tidal volume range:0-1500ml
VENTILATOR
Display Mode High-definition 5.7” LCD screen display
Ventilation Mode IPPV, MANUAL (Manual monitoring of tidal, volume, ventilation volume, respiratory rate).
Ventilation Parameters Tidal volume           50–1500ml

Rate                  2–99bpm

SIMV rate                   2–20bpm

I:E                  2:1–1:8

Inspiratory trigger pressure   -1.0–2.0kPa

Pressure range        1.0–6.0kPa

Parameters for Ventilation

Monitoring

Tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure – time waveform.
SECURITY ALARM SYSTEM
Airway Pressure Alarm Upper limit setting range   1.0–6.0kPa;

Low limit setting range     0.4–2.0kPa.

Per-minute Ventilation Volume Alarm Upper limit setting range   3.0–30L/min;

Low limit setting range     1.0–10L/min.

Sustained High-pressure

Alarm

It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation Alarm It will give sound and light alarm if there is no tidal volume input in 15s
POWER ALARM
Power AC 220V±10%   50Hz±1Hz