Description
Technical Specifications
| PARAMETER | SPECIFICATION | 
| Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant | 
| Scope of Application | Adult, child | 
| Gas Source | O2 0.27–0.55MPa | 
| Flow Meter | O2: 0.1–1.0L/min; 1.1–10L/min | 
| Flow Rate of Rapid Supply | 25–75L/min | 
| Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure < 0.2MPa | 
| Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
 Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.  | 
| Respiratory Circuit | Working mode: all-close, semi-close, semi-open
 Releasing pressure: 0.5-7kPa.  | 
| Respiratory Bellows | Bellows for adults, bellows for children, tidal volume range:0-1500ml | 
| VENTILATOR | |
| Display Mode | High-definition 5.7” LCD screen display | 
| Ventilation Mode | IPPV, MANUAL (Manual monitoring of tidal, volume, ventilation volume, respiratory rate). | 
| Ventilation Parameters | Tidal volume           50–1500ml
 Rate 2–99bpm SIMV rate 2–20bpm I:E 2:1–1:8 Inspiratory trigger pressure -1.0–2.0kPa Pressure range 1.0–6.0kPa  | 
| Parameters for Ventilation
 Monitoring  | 
Tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure – time waveform. | 
| SECURITY ALARM SYSTEM | |
| Airway Pressure Alarm | Upper limit setting range   1.0–6.0kPa;
 Low limit setting range 0.4–2.0kPa.  | 
| Per-minute Ventilation Volume Alarm | Upper limit setting range   3.0–30L/min;
 Low limit setting range 1.0–10L/min.  | 
| Sustained High-pressure
 Alarm  | 
It will give alarm when stress have consistently been higher than 2.5 kPa | 
| Suffocation Alarm | It will give sound and light alarm if there is no tidal volume input in 15s | 
| POWER ALARM | |
| Power | AC 220V±10% 50Hz±1Hz | 