Description
Features
- 8.4″ LCD display
- Temperature monitoring
- Temperature servo control
- Humidity Monitoring
- Humidity Servo Control
- Air temperature and body temperature contorl two modes
- Mattress tilt ±12 °
- Ergonomic designed handle with perfect height
- Trend for 2, 4, 8, 12, 24 hours and up to 7 days
- Front and rear double-walls and air curtain
- Drawers Large (28.0cm x 24.7cm x 36.0cm)
- Drawers Small (28.0cm x 11.3cm x 36.0cm) *2
- Auxiliary Power Output*3
- Backup battery to keep air flow and alarm after power failure for more than 30 mins
Technical Specifications
Performance
Air flow velocity across mattress < 10 cm/sec
Temperature rise time at 22°C (72°F) ambient < 20min (From 22℃, 50%RH, to 35℃)
Temperature variability < 0.5℃
Temperature overshoot < 0.5℃ maximum
Temperature uniformity with a level mattress < 0.8℃
Operating noise level in hood ≤45dBa
Carbon Dioxide (CO2) Level <0.5%
Hood
Tubing access ports: 10
Access door size: 18cm x 13cm
Mattress to hood height: 48.0mm
Soft bed mattress size: 73.6cm x38.6cm x1.8cm
Temperature control modes: Skin and air temperature control mode
Air mode control temperature range: 20℃-37℃
Air mode control override temperature range: 37℃-39℃
Air mode control accuracy ≤±1.0℃
Skin mode control temperature range: 34℃-37℃
Skin mode control override temperature range: 37℃-38℃
Skin temperature accuracy: ±0.3℃
Dual-skin temperature monitoring: Yes
Servo Humidity
Humidity control range: 30%-95% RH
Humidity control operating time without refilling: 24 hours
Humidity control reservoir capacity: 1500ml
Humidity display accuracy: ±5%RH
Humidity control accuracy: ±5% RH
Servo Oxygen (Optional)
Oxygen control range: 21 % to 65%
Oxygen display accuracy of full scale: ±2%
Oxygen control accuracy (100% calibration): ±3%
Oxygen control accuracy (21% calibration): ±5%
Oxygen display resolution: 1%
Scale (Optional)
Weight range: 300g to 8kg
Weight display resolution: 1g
Weight accuracy: ±10g
Operating Environment: 20℃ to 30℃
Humidity: 10 to 95% RH
Air Velocity: Up to 0.3m/sec